< Back to previous page

Publication

Loss and retrieval of a coronary angioplasty stent balloon

Journal Contribution - Journal Article

A case is presented in which a breakage of the coronary stent delivery catheter occurred as retrieval of the stent balloon was attempted after stent implantation. The broken distal balloon shaft with its stuck balloon was retrieved by controlled pulling on the guiding catheter in which a new balloon was inflated, thus trapping the distal shaft. The case underscores the importance of adequate lesion assessment and preparation.
Journal: Cardiovascular Revascularization Medicine
ISSN: 1553-8389
Issue: 4
Volume: 14
Pages: 248-250
Publication year:2013
Keywords:Calcified lesions, Loss of coronary device, Retrieval maneuver
  • Scopus Id: 84881365098