Validity of the Empatica E4 wristband to estimate resting-state heart rate variability in a lab-based context

Tijdschriftbijdrage - Tijdschriftartikel

Lab research might benefit from the advantages of wearable devices, such as their ease of use, to estimate pulse rate (PR) and pulse rate variability (PRV) as an equivalent for heart rate (HR) and heart rate variability. However, before implementing them in a lab context, the validity of the PR and PRV, also on ultra-short time scales (e.g., 30s), needs to be confirmed. We recorded heart activity simultaneously with an E4 wristband and an ECG device in a seated resting condition for 5 min. Our results showed that HR, RMSSD, SDNN and LF, but not HF, were validly estimated by the E4 wristband. Furthermore, the E4 wristband could validly estimate PR with recording lengths as short as 10s. RMSSD and SDNN were validly estimated using 30s or 120 s or an average of multiple short intervals (10s), while HF likely requires longer recording intervals. Based on this study, we formulated several recommendations for using the E4 wristband in a lab context.
Tijdschrift: INTERNATIONAL JOURNAL OF PSYCHOPHYSIOLOGY
ISSN: 0167-8760
Issue: december 2022
Volume: 182
Pagina's: 105 - 118
Jaar van publicatie:2022
Toegankelijkheid:Open