< Terug naar vorige pagina

Publicatie

Non-Invasive Intra-Abdominal Pressure Measurement by Means of Transient Radar Method: In Vitro Validation of a Novel Radar-Based Sensor

Tijdschriftbijdrage - Tijdschriftartikel

Intra-abdominal hypertension, defined as an intra-abdominal pressure (IAP) equal to or above 12 mmHg is one of the major risk-factors for increased morbidity (organ failure) and mortality in critically ill patients. Therefore, IAP monitoring is highly recommended in intensive care unit (ICU) patients to predict development of abdominal compartment syndrome and to provide a better care for patients hospitalized in the ICU. The IAP measurement through the bladder is the actual reference standard advocated by the abdominal compartment society; however, this measurement technique is cumbersome, non-continuous, and carries a potential risk for urinary tract infections and urethral injury. Using microwave reflectometry has been proposed as one of the most promising IAP measurement alternatives. In this study, a novel radar-based method known as transient radar method (TRM) has been used to monitor the IAP in an in vitro model with an advanced abdominal wall phantom. In the second part of the study, further regression analyses have been done to calibrate the TRM system and measure the absolute value of IAP. A correlation of –0.97 with a p-value of 0.0001 was found between the IAP and the reflection response of the abdominal wall phantom. Additionally, a quadratic relation with a bias of −0.06 mmHg was found between IAP obtained from the TRM technique and the IAP values recorded by a pressure gauge. This study showed a promising future for further developing the TRM technique to use it in clinical monitoring.
Tijdschrift: Sensors (Switzerland)
ISSN: 1424-8220
Issue: 18
Volume: 21
Jaar van publicatie:2021
Trefwoorden:intra‐abdominal pressure, non‐invasive measurement, microwave reflectometry, transient radar method
BOF-keylabel:ja
IOF-keylabel:ja
BOF-publication weight:1
Auteurs:International
Authors from:Higher Education, Hospital
Toegankelijkheid:Open