< Terug naar vorige pagina

Publicatie

Mass calibration of the energy axis in ToF-E elastic recoil detection analysis

Tijdschriftbijdrage - Tijdschriftartikel

© 2015 Elsevier B.V. All rights reserved. We report on procedures that we have developed to mass-calibrate the energy axis of ToF-E histograms in elastic recoil detection analysis. The obtained calibration parameters allow one to transform the ToF-E histogram into a calibrated ToF-M histogram.
Tijdschrift: Nuclear Instruments & Methods In Physics Research Section B-Beam Interactions With Materials And Atoms
ISSN: 0168-583X
Volume: 371
Pagina's: 153 - 155
Jaar van publicatie:2016
BOF-keylabel:ja
IOF-keylabel:ja
BOF-publication weight:0.5
CSS-citation score:1
Auteurs:International
Authors from:Government, Higher Education