< Back to previous page

Publication

An ultrasensitive nano UHPLC-ESI-MS/MS method for the quantification of three neuromedin-like peptides in microdialysates

Journal Contribution - Journal Article

AIM: An ultrasensitive nano UHPLC-ESI-MS/MS method is developed to simultaneously monitor three low-concentration neuromedin-like peptides in microdialysates.

RESULTS: Peptide preconcentration and sample desalting is performed online on a trap column. A shallow gradient slope at 300 nl/min on the analytical column maintained at 35°C, followed by two saw-tooth column wash cycles, results in the highest sensitivity and the lowest carryover. The validated method allows the accurate and precise quantification of 0.5 pM neurotensin and neuromedin N (2.5 amol on column), and of 3.0 pM neuromedin B (15.0 amol on column) in in vivo microdialysates without the use of internal standards.

CONCLUSION: The assay is an important tool for elucidating the role of these neuromedin-like peptides in the pathophysiology of neurological disorders.

Journal: Bioanalysis
ISSN: 1757-6180
Issue: 5
Volume: 7
Pages: 605-619
Publication year:2015
Keywords:Chromatography, Liquid, Microdialysis, Neurotensin, Peptide Fragments, Peptides, Tandem Mass Spectrometry
  • VABB Id: c:vabb:395386
  • DOI: https://doi.org/10.4155/bio.14.269
  • Scopus Id: 84926339603
  • WoS Id: 000352294300013
  • ORCID: /0000-0002-1695-7398/work/61423170
  • ORCID: /0000-0003-2140-0751/work/62005212
CSS-citation score:1