< Back to previous page
Publication
Improved analyte detectability of proteins and peptide lysates by means of multiple large-volume injection in LC-MS
Journal Contribution - Journal Article
A 'multiple (trapping) large-volume injection' approach was developed for the analysis of peptides and proteins. In this way, a maximally 10-fold gain in sensitivity could be achieved. The system involves the use of an automated 10-port switching valve in combination with a 1 mm i.d. trapping/guard column and a 1 mm i.d. x 150 mm analytical column. The optimized multiple injection/loading procedure allows quantitative measurements of peptides and protein lysates. Linear calibration curves (R(2) >= 0.988) over a minimum of two orders of magnitude were generated for a range of peptide and protein standards with sensitivities equal to or even exceeding, those generally achieved only through increasing miniaturization (quantification limit >= 0.5 pmol/mL).
Journal: JOURNAL OF SEPARATION SCIENCE
ISSN: 1615-9306
Issue: 14
Volume: 32
Pages: 2346 - 2352
Publication year:2009
Keywords:Analytical, inorganic & nuclear chemistry
Accessibility:Closed