X-ray image quality and system exposure parameters for a hybrid Angio-MR system

Journal Contribution - Journal Article

This study investigates the exposure parameters and required x-ray tube output when performing neurological procedures on a hybrid Angio-MR concept system proposed by Siemens Healthineers. The x-ray part of this system uses a longer source to detector distance than conventional (C-arm) systems and will have a fixed amount of filtration. Additionally, as the x-ray source is situated inside a magnetic field, the focal spot size and shape may be slightly distorted. In order to compare the Angio-MR system to a typical C-arm system, the exposure parameters of 60 thrombectomy procedures, performed in our hospital over the course of one year, were investigated in detail and a set of median values was determined. An analytical simulation platform was then developed to calculate the required tube voltage, tube current and pulse length to reach similar spatial frequency dependent signal difference to noise ratio (SDNR(u)) values as a conventional C-arm angiography system. These simulations were performed for a variety of focal spot sizes for the Angio-MR system. Results show that a standard current x-ray tube has sufficient power to reach similar SDNR(u) values as obtained in a conventional system if the focal spot size between both systems is comparable.