< Back to previous page

Publication

Improved visualisation of real-time recordings during third generation cryoballoon ablation

Journal Contribution - Journal Article

Subtitle:a comparison between the novel short-tip and the second generation device

BACKGROUND: The third-generation Cryoballoon Advance Short-tip (CB-ST) has been designed with a 40 % shortened tip length compared with the former second generation CB advance device (CB-A). Ideally, a shorter tip should permit an improved visualisation of real-time recordings in the pulmonary vein (PV) due to a more proximal positioning of the inner lumen mapping catheter. We sought to compare the incidence of visualisation of real-time recordings in patients having undergone ablation with the CB-ST with patients having received CB-A ablation.

METHODS: All patients having undergone CB ablation using CB-ST technology and the last 500 consecutive patients having undergone CB-A ablation were analysed. Exclusion criteria were the presence of an intracavitary thrombus, uncontrolled heart failure, moderate or severe valvular disease, and contraindications to general anaesthesia.

RESULTS: A total of 600 consecutive patients (58.1 ± 12.9 years, 64 % males) were evaluated (100 CB-ST and 500 CB-A ablations). Real-time recordings were significantly more prevalent in the CB-ST population compared with CB-A group (85.7 vs 67.2 %, p < 0.0001). Real-time recordings could be more frequently visualised in the CB-ST group in all types of veins (LSPV 89 vs 73.4 %, p = 0.0005; LIPV 84 vs 65.6 %, p = 0.0002; RSPV 87 vs 67.4 %, p < 0.0001; RIPV 83 vs 62.4 %, p < 0.0001).

CONCLUSION: The rate of visualisation of real-time recordings is significantly higher during third-generation CB-ST ablation if compared to the second-generation CB-A device. Real-time recordings can be visualised in approximately 85.7 % of veins with this novel cryoballoon.

Journal: Journal of Interventional Cardiac Electrophysiology : An International Journal of Arrhythmias and Pacing
ISSN: 1383-875X
Issue: 3
Volume: 46
Pages: 307-314
Publication year:2016
Keywords:Atrial fibrillation, Catheter ablation, Cryoballoon, Pulmonary vein isolation, Short tip
CSS-citation score:1