NEMA NU 2-2007 performance characteristics of GE Signa integrated PET/MR for different PET isotopes

Journal Contribution - Journal Article

BackgroundFully integrated PET/MR systems are being used frequently in clinical research and routine. National Electrical Manufacturers Association (NEMA) characterization of these systems is generally done with F-18 which is clinically the most relevant PET isotope. However, other PET isotopes, such as Ga-68 and Y-90, are gaining clinical importance as they are of specific interest for oncological applications and for follow-up of Y-90-based radionuclide therapy. These isotopes have a complex decay scheme with a variety of prompt gammas in coincidence. Ga-68 and Y-90 have higher positron energy and, because of the larger positron range, there may be interference with the magnetic field of the MR compared to F-18. Therefore, it is relevant to determine the performance of PET/MR for these clinically relevant and commercially available isotopes.MethodsNEMA NU 2-2007 performance measurements were performed for characterizing the spatial resolution, sensitivity, image quality, and the accuracy of attenuation and scatter corrections for F-18, Ga-68, and Y-90. Scatter fraction and noise equivalent count rate (NECR) tests were performed using F-18 and Ga-68. All phantom data were acquired on the GE Signa integrated PET/MR system, installed in UZ Leuven, Belgium.Results(18)F, Ga-68, and Y-90 NEMA performance tests resulted in substantially different system characteristics. In comparison with F-18, the spatial resolution is about 1mm larger in the axial direction for Ga-68 and no significative effect was found for Y-90. The impact of this lower resolution is also visible in the recovery coefficients of the smallest spheres of Ga-68 in image quality measurements, where clearly lower values are obtained. For Y-90, the low number of counts leads to a large variability in the image quality measurements. The primary factor for the sensitivity change is the scale factor related to the positron emission fraction. There is also an impact on the peak NECR, which is lower for Ga-68 than for F-18 and appears at higher activities.ConclusionsThe system performance of GE Signa integrated PET/MR was substantially different, in terms of NEMA spatial resolution, image quality, and NECR for Ga-68 and Y-90 compared to F-18. But these differences are compensated by the PET/MR scanner technologies and reconstructions methods.
Journal: EJNMMI PHYSICS
ISSN: 2197-7364
Volume: 6
Publication year:2019
BOF-keylabel:yes
IOF-keylabel:yes
BOF-publication weight:2
CSS-citation score:2
Authors:International
Authors from:Higher Education
Accessibility:Open