Characterisation and quenching correction for an Al2O3:C optical fibre real time system in therapeutic proton, helium, and carbon-charged beams

Journal Contribution - Journal Article

Real time radioluminescence fibre-based detectors were investigated for application in proton, helium, and carbon therapy dosimetry. The Al2O3:C probes are made of one single crystal (1 mm) and two droplets of micro powder in two sizes (38 mu m and 4 mu m) mixed with a wate-requivalent binder. The fibres were irradiated behind different thicknesses of solid slabs, and the Bragg curves presented a quenching effect attributed to the nonlinear response of the radioluminescence (RL) signal as a function of linear energy transfer (LET). Experimental data and Monte Carlo simulations were utilised to acquire a quenching correction method, adapted from Birks' formulation, to restore the linear dose-response for particle therapy beams. The method for quenching correction was applied and yielded the best results for the '4 mu m' optical fibre probe, with an agreement at the Bragg peak of 1.4% (160 MeV), and 1.5% (230 MeV) for proton-charged particles; 2.4% (150 MeV/u) for helium-charged particles and of 4.8% (290 MeV/u) and 2.9% (400 MeV/u) for the carbon-charged particles. The most substantial deviations for the '4 mu m' optical fibre probe were found at the falloff regions, with similar to 3% (protons), similar to 5% (helium) and 6% (carbon).
Journal: SENSORS
ISSN: 1424-8220
Issue: 23
Volume: 22
Publication year:2022
Accessibility:Open